3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 5.2 Tube Detect Sensor (Beam Block) Adjustment
3 - 73
3
6. [NON-CE MARK ONLY] - Adjust Gain (
a. Attach Resistor tool between TP22 and GND (TP23).
b. Set volt meter to AC volts.
c. Connect positive (+) lead to R12 at end near RA14. Connect negative (-) lead to GND (TP23)