CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 12.1 Blood In-Line Sensor Check/Adjustment
3 - 151
3
5. Place 0.80 M/µl sample in position 1 of rack 1 (
Figure 3-80 Dilution Location
6. Minimum Volume - 1.0 ml.
7. Connect DC meter to Blood In-Line Sensor board TP4 (+) to TP3 (-) GND.
SL-106
0.8 m/ul
Dilution