CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 6.5 Mixing Limit Sensor Check/Adjustment
3 - 103
3
VP - 6.5 Mixing Limit Sensor Check/Adjustment
Purpose: Perform this procedure to verify and Adjust Mix Assembly Limit Sensor.
E Loader
[CE MARK]
1. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
2. Clear all racks from under mixing paddle.
3. Press (Z
↓
) key down until mixing paddle stops.
4. Check that distance between platen to knurled edge of mixing paddle is 3.27 in. (83mm) (
5. If check fails, adjust Limit Sensor.
Special Tools
DESCRIPTION
QTY
PN
Rack Guide
Alignment Tool
1
9420084
Mixing Head
Alignment Tool
1
9420082