4
5
1
1
INSTRUCTIONS FOR USE – Wells, Buttons, & Display Interpretation
Please familiarize yourself with the face of the 3M™ Attest™ Auto-reader 390G:
Wells
1. Incubation/Reader Wells – The 3M™ Attest™ Auto-reader 390G has ten
incubation wells.
Buttons
2. Information / Up Arrow Button
3. Temperature / Down Arrow Button
4. Buzzer Alarm OFF / Enter Button
Information Display
5. LCD Display – Display is backlit
6. LCD Display – Example of Error Message for well #6
7. LCD Display – Example of Remaining Incubation Time, in minutes, for well #10
8. LCD Display – Example of Negative ( - ) BI result for well #9
9. LCD Display – Example of Positive ( + ) BI result for well #8
10. LCD Display – Alarm Enabled Indicator
Color-Coded Configuration Indicator
11. Incubation well configuration sticker is color-coded green
BI Crusher
12. BI Crusher Well
INSTRUCTIONS FOR USE – 3M™ Attest™
Auto-reader 390G Configuration
Note:
Ideally, the 3M™ Attest™ Auto-reader 390G should be empty (i.e., have no
incubating BIs) when it is reconfigured.
Press the
and
buttons on the front of the 3M™ Attest™ Auto-reader 390G
simultaneously for 3 seconds to enter configuration mode. Once in configuration mode,
use the Up
and
Down buttons to set the 3M™ Attest™ Auto-reader 390G
clock and audible alarm as desired. The
button acts as an “Enter” key and moves the
flashing cursor to the next changeable setting.
Setting the 3M™ Attest™ Auto-reader 390G Clock
The 3M™ Attest™ Auto-reader 390G date and time are displayed in this format:
YYYY-MM-DD HH:MM, with the time displayed using a 24 hour clock. To change the year,
if necessary, push the up or down buttons to set desired year, then press
to advance
to the month, adjust as necessary and press
to advance to the day. Use the
button to move to the time and set the hours and minutes.
Setting the Audible Alarm Function
Move across the top row of text by pushing the
button. The flashing “B” in Buzzer
indicates this is a changeable setting. It is recommended this be set at “Buzzer Enabled”
to audibly alert the operator to a positive BI but it may be set to “Buzzer Disabled” by
using either the up or down button. Note: If Buzzer Disabled is selected, and the changes
accepted, the speaker icon ( ) will disappear from the LCD panel and the alarm will be
disabled for all positive 3M™ Attest™ Rapid Readout Biological Indicators, error codes,
and caution code alarm warnings. Note: The 3M™ Attest™ Auto-reader 390G display
backlight flashes whenever an audible alarm sounds. Disabling the audible alarm does
NOT disable the LCD display backlight flashing.
Press
to select the desired option and advance to the Accept screen.
Accept Screen
Use the up or down buttons to toggle between Yes and No, press
to accept changes
(Yes selected) or retain current settings (No selected) and exit the configuration mode. The
LCD panel will return to the standard display.
SPECIFICATIONS
Power Supply Specifications
Power supply
Operating Condition
Units
Voltage
Range
100-240
Volts AC
Frequency
50/60
Hertz
Current
0.6
Amps
Environmental Operating Conditions
Environmental Condition
Operating Condition
Units
Altitude
3000 (max)
Meters
Operating
Temperature
16 –
29
ºC
Storage
Temperature
-10 to
50
ºC
Operating Relative Humidity
20-80 (non-condensing) %
Voltage
Range
12
Volts DC
Frequency
DC
Current
1.2
Amps
Installation/Over
Voltage
Category II
Pollution
Degree
2
INSTRUCTIONS FOR USE – Initial Set-up
Choose Location
Place 3M™ Attest™ Auto-reader 390G on a firm and level surface, away from sunlight
and incandescent light. To save counter space, units may be stacked two high.
WARNING:
• Do not place the instrument in environment exposed to sunlight or strong
incandescent light.
• Do not place instrument close to any device that emits a strong
electro-magnetic field.
• Do not use on a vibrating surface.
Select & Install Correct Power Plug
The 3M™ Attest™ Auto-reader 390G power supply comes with multiple blade
configurations. Install the blade configuration appropriate for your country into the power
supply and plug into appropriate AC grounded outlet.
WARNING:
Use only the power supply specified for this product and certified for
the country of use.
Wait 30 Minutes Before Using
Once plugged in, allow a 30 minute warm-up period before placing 3M™ Attest™ Rapid Readout Biological Indicators into the incubation wells. The C1 caution code will disappear
when proper incubation temperature is reached.
Special Considerations
It is recommended that the unit be left on to eliminate warm-up periods.
The unit may be turned off when it is anticipated that it will not be used for an extended period.
Rendered on 9/12/2013 10:34:26 AM
Page 4