Harmonic emissions
IEC 61000-3-2
NA
The Endo-Eze
TM
Find® does
not have AC power input or AC
power output, therefore testing
for this is not applicable.
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
NA
(IEC 60601-1-2)
Table 2: Guidance and Manufacture’s Declaration for
Electromagnetic Immunity
The Endo-Eze
TM
Find® is intended for use in the electromagnetic environment
specified below. The customer or user should ensure that it is used in such
an environment.
IMMUNITY test
IEC 60601
test level
Compliance
level
Electromagnetic
environment guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6kV
contact
± 8 kV air
± 6kV contact
± 8 kV air
The Endo-Eze
TM
Find® has
built in ESD protection.
The physical environment
should be restricted to the
following:
1. Humidity range: 10%
- 90%
2. Temperatures range:
10
°
C - 42
°
C
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV
for power
supply
lines
± 1 kV
for input/
output
lines
Not
applicable
Not
applicable
The Endo-Eze
TM
Find®
does not have AC power
input or AC power output,
therefore testing for
this is not applicable.
Interconnecting cables
are not longer than 3m,
therefore testing is not
applicable.
Surge
IEC 61000-4-5
± 1 kV line
to line
± 2 kV line
to earth
Not
applicable
Not
applicable
Voltage
dips, shorts,
interruptions and
variations on the
power supply
input lines
IEC 61000-4-11
<5% U
(>95% dip
in U for 0.5
cycle)
40% U
(60% dip
in U for 5
cycles)
70% U
(30% dip
in U for 25
cycles)
<5% U
(>95% dip
in U for
5 s)
Not
applicable
Not
applicable
Not
applicable
Not
applicable
The Endo-Eze
TM
Find®
does not have AC power
input or AC power output,
therefore testing for this is
not applicable.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical,
residential, commercial,
hospital, or military
environment.
NOTE:
U
is the AC. mains voltage prior to application of the test level
Note 1:
The Endo-Eze
TM
Find® is not equipped with any ports or any
accessible I/O lines.
Note 2:
The Endo-Eze
TM
Find® may be susceptible to strong magnetic fields
or static electric fields higher than the limits specified in testing,
which could interrupt normal operation. If there is a reason to
believe that this type of interference has occurred, simply remove
and re-insert the batteries and the unit should reset.
12. Disclaimer
The manufacturer, its representatives and its dealers shall
have no liability or responsibility to customers or any other
person or entity with respect to any liability, loss or damage
caused or alleged to be caused directly or indirectly by
equipment sold or furnished by us, including, but not
limited to, any interruption of service, loss of business or
anticipatory profits, or consequential damages resulting
from the use or operation of the equipment.
The manufacturer reserves the right to implement changes
and modifications of the product at any time, to revise this
publication and to make changes in the contents hereof
without obligation to notify any person of such changes,
modifications or revisions.
13. Technical Specifications
The Endo-Eze
TM
Find® belongs to the following category of
medical devices:
- Internally powered equipment (AAA 1.5V alkaline
battery)
- Type BF applied parts
- Not suitable for use in presence of flammable anesthetic
mixtures with air, oxygen or nitrous oxide
- Continuous operation
- Ingress of liquids – not protected
- The device is intended for indoor use only
- Environmental conditions during transportation:
temperature: –20
�C to +60�
C
(0 to 140
�F); relative humidity: 10% to 90%, non-
condensing
Endo-Eze
TM
Find® is intended for use in electromagnetic
environment specified for equipment of Group 1 Class B.
Specifications:
Dimensions: W55 x H92 x T16 mm
Weight: 100 gr.
Type of screen: Custom Color Graphic LCD
Screen dimensions: 51 x 38 mm
Supply: AAA 1.5V alkaline battery
14. Electromagnetic declarations
(IEC 60601-1-2)
Table 1:
Guidance and Manufacture’s Declaration for
Electromagnetic Emissions
The Endo-Eze
TM
Find® is intended for use in the electromagnetic environment
specified below. The customer or user should ensure that it is used in such
an environment.
Emissions Test
Compliance
Electromagnetic environment
- guidance
RF emissions
EN 5011
(European Norm
equivalent to CISPR 11)
Group 1
The Endo-Eze
TM
Find® uses
low voltage and low current,
electrical and electromagnetic
energy only for its internal
functions. Therefore, its RF
emissions are very low and are
not likely to cause interference
in nearby electronic
equipment.
RF emission
EN 5011
(European Norm
equivalent to CISPR 11)
Class B