5
or larger files may be used when anatomically necessary.
• For convenience of use it is recommended that selected
file will be about 5mm longer than the radiographic
working length.
• Before measuring, check the accessories and verify
good connectivity.
• Endo-Eze
TM
Find® should not be dropped or subjected to
other manners of impact or shock as this could result in
malfunction of the device and voiding of the warranty.
• When disconnecting the lip clip, file clip or measuring
cable, always grip the connectors.
Do not
pull directly
on the wires to disconnect.
• Discard any old, damaged or leaky battery. Battery
should be disposed of according to local codes and
regulations. Battery cannot be disposed of as municipal
waste within the European Union.
• Ensure the battery is installed with correct polarity.
•
Do not immerse Endo-Eze
TM
Find® unit. This may
damage the device and will void the warranty.
• Only use recommended cleaning solutions on
the device. See the
“Maintenance, cleaning and
sterilization”
section 7 (page 7) for recommended
cleaning and sterilization.
• The items which may contact the patient during
intended use of the device (lip clip, file clip and touch
probe) must be cleaned and sterilized before the first
use and between patients to avoid the risk of cross-
contamination.
• Care should be exercised when the Endo-Eze
TM
Find®
is used around devices emitting electromagnetic noise
such as fluorescent lamps, film viewers, ultrasonic
devices, radio frequency transmitters, cell phones,
remote controls, or devices that radiate electromagnetic
(EM) fields, areas where there are strong magnetic
fields, or conditions where there is high static electricity
(ESD). While the Endo-Eze
TM
Find® has been built to
withstand these conditions, EM and ESD fields may
interfere with the device’s proper operation. If the
Endo-Eze
TM
Find® must be used under these conditions,
take care to see that the device is operating normally
(See manufacturer’s declaration of electromagnetic
compatibility).
4. Package Contents
Check the contents of the package before use:
• 1 Endo-Eze
TM
Find®
•
1 Cradle
• 1 AAA 1.5V Alkaline Battery
•
1 Measuring cable
•
2 File clips
•
5 Lip clips
•
1 Touch probe
• 1 Screwdriver (for battery compartment)
•
1 User Manual
5. Getting Started
5.1. General
There are two buttons on the front panel:
On/Off
Volume adjustment
The measuring cable receptacle is located on the left side
of the device.
The battery compartment is located on back of the device.
The Endo-Eze
TM
Find® may be placed in or out of the cradle.
The lip clip, the touch probe and the file clip should be
sterilized before use. Please refer to “
Maintenance
,
Cleaning and Sterilization”
section 7 (page 7) for further
information regarding maintenance of the Endo-Eze
TM
Find®.
5.2. Installing / Replacing the Battery
The Endo-Eze
TM
Find® is powered by a standard AAA size
1.5V alkaline battery (included).
5.2.1. To install/replace the battery, unscrew and
remove the battery compartment cover on
the back of the device (Fig. 1a). Remove the
old battery (if one is present) using the built-
in cloth strip; insert the new battery following
polarity indications marked inside the battery
compartment, (Fig. 1b), replace cover and tighten
screw.
1 - Measuring cable
2 – File clips
5 - Lip clips
1 – Touch Probe
Fig. 1a
Fig. 1b
English