3152B User Manual
Publication No. 980935 Rev. A
3152B output at the DMM input with the 50
Ω
feedthrough
termination
3. Configure the 3152B as follows:
Amplitude:
20
mV
Offset
+3
V
Output:
On
Adjustment:
1. CAL:SETUP 47 for a DMM reading of +3 V
±
15 mV
Setup 20
+5 V Offset Output, Amplifier In
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
10
V
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination
3. Configure the 3152B as follows:
Amplitude:
20
mV
Offset
+5
V
Output:
On
Adjustment:
1. CAL:SETUP 48 for a DMM reading of +5 V
±
25 mV
Setup 21
+7 V Offset Output, Amplifier In
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
10
V
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination
3. Configure the 3152B as follows:
Amplitude:
20
mV
Offset
+7
V
Output:
On
Adjustment:
1. CAL:SETUP 49 for a DMM reading of +7 V
±
35 mV
Setup 22
-1 V Offset Output, Amplifier In
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
7-14 Adjustments and Firmware Update
EADS North America Test and Services
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com