Publication No. 980935 Rev. A
3152B User Manual
EADS North America Test and Services
Programming Reference 5-15
Table 5-2, 3152B SCPI Command Summary (continued)
Keyword
Parameter Form
Default
3152A
HS
Output Control Commands
:OUTPut
9
:SHUNt
OFF | ON | 0 | 1
Error, not
supported
9
0250A
:ECLTrg<n>
<n> = 0 to 1
0
[:STATe]
OFF | ON | 0 | 1
0,0
9
0210B
:FILTer
9
[:LPASs]
9
:FREQuency
20MHz | 25MHz | 50MHz | 60MHz | 120 MHz
(20MHz/25MHz = 25M)
20M
9
0201A
[:STATe]
OFF | ON | 0 | 1
0
9
0202A
:LOAD
50 to 1e6
50
0012A
[:STATe]
OFF | ON | 0 | 1
0
9
0200A
:SYNC
9
[:STATe]
OFF | ON | 0 | 1
0
9
0241A
:POSition
9
[:POINt]
0 to 1e6-1 (0 to 2/4e6-1 with option 1/2)
0
9
0242A
:SOURce
BIT | LCOMplete | SSYNc | HCLock | PULSe | ZEROcross
BIT
9
0240A
:WIDTh
4
to
n-8
4
9
0243A
:TRIGger
9
:SOURce
BIT
|
LCOMPlete
| INTernal | EXTernal
BIT
9
0220A
:TTLTrg<n>
<n> = 0 to 7
0
9
[:STATe]
OFF | ON | 0 | 1
0,0
9
0230B
[:SOURce]
:ROSCillator
:SOURce
INTernal
|
EXTernal | CLK10
INT/CLK10
0028A
:FREQuency
9
:EXTernal?
9
0420@?
[:CW]
10e-3 to 100e6 | MINimum | MAXimum
1e6
9
0400A
:RASTer
10e-6 to 250e6 | MINimum | MAXimum
1e6
9
0410A
:SOURce
INTernal | EXTernal | ECLTrg0 | LBUS
INT
9
0411A
:VOLTage
9
[:LEVel]
9
[:AMPLitude]
10e-3 to 16 | MINimum | MAXimum
5
9
0500A
:OFFSet
-7.992
to
7.992
0
9
0501A
:PHASe
[:OFFSet]
0 to 1e6-1 (0 to 2e6-1 with option 2)
0
0027A
:FUNCtion
:MODE
FIXed | USER | SEQuence | | SWEep (calculated)(*) |
MODulation | DPULse | HALFcycle | COUNter
FIX
9
0610A
(*) Legacy Mode defaults to CLK10; Modern Mode defaults to INT