Publication No. 980935 Rev. A
3152B User Manual
1. CAL:SETUP 11 for a DMM reading of -1 V
±
5 mV; note
reading.
2. Repeat Setup 26 and Setup 27 until errors are balanced
between the steps.
Amplitude
Adjustments
The amplitude adjustments ensure that the AC levels are within the
specified range. Use this procedure if you suspect that amplitude
accuracy is an issue.
Setup 28
10 V Amplitude, Amplifier In - Arbitrary
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
10
V
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination
3. Configure the 3152B as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
10
V
Adjustment:
1. Adjust CAL:SETUP12 for a DMM reading of 3.535 V
±
35 mV
Setup 29
5 V Amplitude, Amplifier In - Arbitrary
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
10
V
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination.
3. Configure the 3152B as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
5
V
Adjustment:
1. Adjust CAL:SETUP 13 for a DMM reading of 1.767 V
±
20 mV
EADS North America Test and Services
Adjustments and Firmware Update 7-17