TRIangle:PHASe <phase>(?)
HS Syntax: 0710A<float>(?)
Description
Programs the start phase for the standard triangular waveform. This command has no affect on modulated
waveforms.
Parameters
Name
Range
Type
Default
Description
<phase>
0 to 360
Numeric
0
Programs the start phase parameter in units of degrees.
Response
The 3152B returns the current start phase value for the triangle function.
TRIangle:POWer <power>(?)
HS Syntax: 0711A<integer>(?)
Description
Programs power for the triangle
x
waveform. This command has no affect on modulated waveforms.
Parameters
Name
Range
Type
Default
Description
<power>
1 to 9
Numeric
1
Programs the power that the triangle function will be
raised to. Note that even powers will provide waveforms
that are positive (all points >= 0) whereas odd powers
will provide functions that are positive or negative 50%
of the time.
Response
The 3152B returns the current power value for the triangle
x
function.
SQUare:DCYCle <duty_cycle>(?)
HS Syntax: 0720A<float>(?)
Description
Programs duty cycle of the standard square waveform.
Parameters
Name
Range
Type
Default
Description
<duty_cycle>
0 to 99.99 Numeric
50
Programs the square wave duty cycle parameter in
units of percent (ratio of on-time to off-time).
Response
The 3152B returns the current duty cycle value for the square wave function.