3152B User Manual
Publication No. 980935 Rev. A
Test Results
Pass
Fail
SYNC Output - HCL
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
As required by the test
Amplitude:
2 V/div
2. Connect the 3152B output to the oscilloscope input (1)
3. Connect the 3152B SYNC to the oscilloscope input (2)
4. Configure model 3152B channel as follows:
Waveform: Sine
Output: On
Sync Output:
On
Sync Source: HCL
Test Procedure:
1. Observe that the SYNC output has a 50% duty cycle.
Test Results
Pass
Fail
SYNC Output - Pulse
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
As required by the test
Amplitude:
2 V/div
2. Connect the 3152B output to the oscilloscope input (1)
3. Connect the 3152B SYNC to the oscilloscope input (2)
4. Configure model 3152B channel as follows:
Waveform: Sine
Output: On
Sync Output:
On
Sync Source: Pulse
Sync Width:
8
Sync Position: 48
Test Procedure:
1. Observe that the SYNC output has a width of 8 sample clock
cycles and that its position is shifted by 48 sample clock
periods.
Test Results
Pass
Fail
6-20 Performance Checks