Publication No. 980935 Rev. A
3152B User Manual
Setup 17
Base Line Offset, Amplifier Out - Arbitrary
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
100
mV
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination
3. Configure the 3152B as follows:
Output:
On
Amplitude:
1
V
Adjustment:
1. Adjust CAL:SETUP 7 for a DMM reading of 0 V
±
5 mV
Offset
Adjustments
The offset adjustments ensure that the DC offsets are within the
specified range. Use this procedure if you suspect that offset accuracy
is an issue.
Setup 18
+ 1 V Offset Output, Amplifier In
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
1
V
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination
3. Configure the 3152B as follows:
Amplitude:
20
mV
Offset
+1
V
Output:
On
Adjustment:
1. CAL:SETUP 46 for a DMM reading of +1 V
±
5 mV
Setup 19
+3 V Offset Output, Amplifier In
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
10
V
2. Connect the 3152B output to the DMM input. Terminate the
EADS North America Test and Services
Adjustments and Firmware Update 7-13
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com