Publication No. 980935 Rev. A
3152B User Manual
EADS North America Test and Services
Programming Reference 5-97
Response
The 3152B returns the current pulse period value in units of seconds.
DPULse:TRANsition <rise>(?)
HS Syntax: 0151A<float>(?)
Description
Programs the interval it will take the pulse to transition from its low to high level settings. The parameter is
programmed in units of seconds.
Parameters
Name
Range
Type
Default
Description
<rise>
0 to 1e3
Numeric
1e-3
Sets the rise time parameter. Note that the sum of all
parameters, including the rise time must not exceed the
programmed pulse period and therefore, it is
recommended that the pulse period be programmed
before all other pulse parameters.
Response
The 3152B returns the current rise time value in units of seconds.
DPULse:TRANsition:TRAiling <fall>(?)
HS Syntax: 0152A<float>(?)
Description
Programs the interval it will take the pulse to transition from its high to low level settings. The parameter is
programmed in units of seconds.
Parameters
Name
Range
Type
Default
Description
<fall>
0 to 1e3
Numeric
1e-3
Sets the fall time parameter. Note that the sum of all
parameters, including the fall time must not exceed the
programmed pulse period and therefore, it is
recommended that the pulse period be programmed
before all other pulse parameters.
Response
The 3152B returns the current fall time value in units of seconds.