3152B User Manual
Publication No. 980935 Rev. A
Range:
100
mV
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination
3. Configure the 3152B as follows:
Mode:
Modulation
Output:
On
Amplitude:
6
V
Adjustment:
1. Adjust CAL:SETUP 4 for a DMM reading of 0 V
±
20 mV.
Setup 15
Base Line Offset, Amplifier Out – Modulation
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
100
mV
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination
3. Configure the 3152B as follows:
Mode:
Modulation
Output:
On
Amplitude:
1
V
Adjustment::
1. Adjust CAL:SETUP 5 for a DMM reading of 0 V
±
5 mV
Setup 16
Base Line Offset, Amplifier In – Arbitrary
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
100
mV
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination
3. Configure the 3152B as follows:
Output:
On
Amplitude:
6
V
Adjustment:
1. Adjust CAL:SETUP 6 for a DMM reading of 0 V
±
20 mV
7-12 Adjustments and Firmware Update
EADS North America Test and Services