3152B User Manual
Publication No. 980935 Rev. A
3. Using ArbConnection prepare and download the following
waveform:
Wavelength: 100
points
Waveform:
Pulse, Delay = 0.1, Rise/Fall = 0, High Time
= 99.9
4. Configure the 3152B as follows:
SCLK 100
MS/s
Waveform: Arbitrary
Amplitude: 5
V
Run Mode:
Triggered
Trigger Level 0 V
Re-trigger: On
Re-trigger Delay: As required by the test
Trigger Source: Bus
Output: On
Test Procedure
1. Manually trigger the instrument.
2. Perform trigger delay tests using Table 6-23.
Table 6-23, Re-Trigger Delay Tests
3152B Re-trigger
Setting
Error Limits
Counter Reading
Pass
Fail
1
μ
s 1
μ
s
±
85 ns
1 ms
1 ms
±
50
μ
s
1 s
1 s
±
50 ms
10 s
10 s
±
500 ms
Trigger Slope
Equipment: Oscilloscope, function generator
Preparation:
1. Configure the Oscilloscope follows:
Termination: 50
Ω
, 20 dB feedthrough attenuator at the
oscilloscope input
Setup:
As required for the test
Trigger Source: External
2. Connect 3152B output to the oscilloscope input
3. Configure the function generator as follows:
Frequency 10
kHz
Run Mode:
Continue
Waveform:
2 V Square
4. Connect the function generator TTL output to the 3152B TRIG
IN connector
5. Connect the function generator main output to the 2
nd
channel
of the oscilloscope
6. Configure the 3152B as follows:
Frequency: 1
MHz
Waveform: Sine
wave
Run Mode:
Triggered
6-14 Performance Checks