3152B User Manual
Publication No. 980935 Rev. A
3. Connect the 3152B SYNC to the oscilloscope input, channel 2.
4. Configure model 3152B controls as follows:
Waveform:
Modulated
Modulation:
AM
Carrier Freq:
1 MHz
Mod Frequency: 1 kHz
Mod Depth:
50 %
Mod Wave
Sine
Sync: On
Output: On
Test Procedure:
1. Verify AM operation on the oscilloscope as follows:
Waveform:
Amplitude modulated sine
Mod depth:
50 %
±
5 %
Test Results
Pass
Fail
FSK
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
0.2 ms
Sampling Rate: 50 MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude: 1
V/div.
2. Connect the 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel 2.
4. Configure model 3152B controls as follows:
Waveform:
Modulated
Modulation:
FSK
Carrier Freq:
2 MHz
Shift Frequency: 4 MHz
Baud Rate:
10 kHz
Marker Index: 1
Sync: On
Output: On
5. Using ArbConnection, prepare and download 10-step FSK list
with alternating “0” and “1”
Test Procedure:
1. Verify FSK operation on the oscilloscope as follows:
Waveform: Square
wave
Period: 0.2
ms
Max Freq.:
4 MHz
Min Freq.:
2 MHz
Test Results
Pass
Fail
6-30 Performance Checks