Publication No. 980935 Rev. A
3152B User Manual
Table 6-4, Amplitude Accuracy, DAC output
3152B Amplitude
Setting
Error Limits
DMM Reading
Pass
Fail
10.00 V
3.534 V
±
85 mV
5.000 V
1.767 V
±
67 mV
1.000 V
353.4 mV
±
23 mV
500m V
176.7 mV
±
22 mV
100.0m V
35.34 mV
±
5.4 mV
50.00m V
17.67 mV
±
5.2 mV
Amplitude Accuracy,
DDS Output
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Termination: 50
Ω
feedthrough at the DMM input
Function: ACV
2. Connect 3152B output to the DMM input
3. Configure the 3152B as follows:
Waveform: Modulated
Modulation:
OFF
CW Frequency: 1 kHz
Output: On
Amplitude:
As specified in Table 6-5
Test Procedure
1. Perform amplitude Accuracy tests using Table 6-5.
Table 6-5, Amplitude Accuracy, DDS output
3152B Amplitude
Setting
Error Limits
DMM Reading
Pass
Fail
10.00 V
3.534 V
±
85 mV
5.000 V
1.767 V
±
67 mV
1.000 V
353.4 mV
±
23 mV
500m V
176.7 mV
±
22 mV
100.0m V
35.34 mV
±
5.4 mV
50.00m V
17.67 mV
±
5.2 mV
Offset Accuracy
The offset accuracy test checks the accuracy of the offset generator.
The offset path is checked for both the DAC route (arbitrary and
standard waveforms) and the DDS route (CW and modulated
waveforms).
Offset Accuracy,
Equipment: DMM