Publication No. 980935 Rev. A
3152B User Manual
EADS North America Test and Services
Programming Reference 5-123
or activating sequence mode when the 3152B is set to burst
mode. Corrective action: Remove the 3152B from burst or
sequence and then selected the desired mode.
5. Changing operating mode from triggered to continuous when the
3152B is set to single sequence advance, or changing the
operating mode from continuous to triggered when the 3152B is
set to automatic sequence advance mode. Corrective action:
Observe the 3152B advance mode while setting sequence
advance.
There are other setting conflict errors, which are exclusively dedicated
for the pulse function. These errors are listed and described in
Chapter 3, under the pulse function description
-222,”Data out of range”. Parameter data, which followed a specific
header, could not be used because its value is outside the valid range
defined by the generator.
-224,”Illegal parameter value”. A discrete parameter was received
which was not a valid choice for the command. An invalid parameter
choice may have been used.
-300,”Device-specific-error”. This is the generic device-dependent
error for the instrument when it cannot detect more specific errors. A
device- specific error as defined in IEEE-488.2 has occurred.
-311,”Memory error”. Indicates that an error was detected in the
instrument’s memory.
-350,”Queue Overflow”. The error queue is full because more than 30
errors have occurred. No additional errors are stored until the errors
from the queue are removed. The error queue is cleared when power
has been shut off, or after a *CLS command has been executed.
-410,”Query INTERRUPTED”. A command was received which sends
data to the output buffer, but the output buffer contained data from a
previous command (the previous data is not overwritten). The output
buffer is cleared when power is shut off or after a device clear has
been executed.