Publication No. 980935 Rev. A
3152B User Manual
EADS North America Test and Services
Programming Reference 5-27
Response
The 3152B will return ADJ, ECLT or LBUS depending on the current instrument couple path assignment.
INSTrument:COUPle:SLAVe:DELete <LAN_address>
HS Syntax: 00040<string>(?)
Description
This command will delete a designated slave instrument from a synchronized multi-instrument system list.
This command is associated with LAN operation only.
Parameters
Name
Type
Default
Description
<LAN_address>
String
Will remove a designated instrument (specified by IP
address) from a synchronized multi-instrument system
list. Contact your IT department, if you are not sure how
to connect to a LAN address.
INSTrument:COUPle:SLAVe:INSert <3152B>,<LAN_address>
HS Syntax: 00051<3152B,string>(?)
Description
This command will add a designated slave instrument to a synchronized multi-instrument system list. This
command is associated with LAN operation only.
Parameters
Name
Type
Default
Description
<3152B>, <LAN_address> String
Specifying the correct model number is crucial for
correct assignment of the instrument designator and for
selection of the correct instrument number for the
INST:SEL command.
INSTrument:COUPle:STATe {OFF|ON|0|1}(?)
HS Syntax: 0006A<0|1>(?)
Description
This command will turn the 3152B coupled state on and off.
Parameters
Range
Type
Default
Description
0-1
Discrete
0
Turns coupled mode on and off. Note that this command
must be applied to the master instrument only, otherwise,
slave instruments will not couple. To select the master
instrument use the INST:SEL 1 command.
Response
The 3152B returns 1 if the coupled state is on or 0 if the couple state is off.