Minimum Segment Size in a Sequence
10 points
Custom Sequence Creation Software
ArbConnection software allows instrument control
and creation of custom sequences. Sequences are
built as tables and downloaded to the instrument
Download Modes
Normal, using SCPI commands; Fast, using binary
data (<250 ms / 100 steps)
Modulated Waveform Generator Characteristics
General
Description
Using this mode of operation, one may select from a
list of built-in modulation schemes.
Carrier Waveform (CW)
Sinewave
Modulation Source
Internal
Run Modes
Off (outputs CW), Continuous, Triggered, Delayed
Trigger, Re-trigger, Burst and Gated
Interrupted Modulation Carrier Idle Mode
On or Off, programmable
Run Mode Advance Source
Front panel TRIG IN, TTLTRG(0-7), Internal trigger,
Software commands
Trigger Delay
(Trigger input to modulation output)
[(0; <100 ns to 20 s) + system delay]
Resolution
20
ns
Error
6 sample clock 150 ns
Re-trigger Delay
(Modulation end to modulation restart)
<
100 ns to 20 s
Resolution
20
ns
Error
3 sample clock 20 ns
Trigger Parameters
All trigger parameters such as level, slope, jitter, etc.
apply
Marker Output
Description
Marks the crossing of a specific frequency and step
setting. The marker pulse is generated through the
SYNC connector. Marker placement is available for
all modulation modes except AM
Output
SYNC connector, TTLTRG(0-7)
Level TTL
Marker Type
Single pulse at the specified frequency
Position
Programmable for a specific frequency setting
Sweep
Swept Waveform
Sine, square or triangle
Sweep Step
Linear or log
Sweep Direction
Up or Down
Sweep Range
10 Hz to 100 MHz
Sweep Time
1.4
μ
s to 40 s
FM
Modulated Waveform
Sine wave
Modulating Waveforms
Sine, square, triangle, Ramp
Carrier Frequency Range
10 Hz to 100 MHz
Modulating Frequency Range
10 mHz to 350 kHz
A-6 3152B Specifications