3152B User Manual
Publication No. 980935 Rev. A
3-36
Operation
EADS North America Test and Services
valid trigger signal, and ends after the gate has closed. Processing
time for the reading and the display is roughly 100 ms. In this mode,
the counter can take a maximum of ten readings per second.
2. Gate time period must be higher than the signal period
The gate must open for an interval that allows enough transitions to
pass through the counter gate. If the gate time is too short to
measure a signal, the gate will open, but no results can be
obtained.
3. Auxiliary functions disable waveform generation
When the Counter/Timer mode is selected, all operations of the
waveform generator are stopped, and the waveform memory is
purged.