Publication No. 980935 Rev. A
3152B User Manual
Adjustment:
1. Adjust CAL:SETUP 21 for a DMM reading of 17.67 mV
±
0.5 mV
Setup 38
3 V Amplitude, Post Attenuators - Arbitrary
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
10
V
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination.
3. Configure the 3152B as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
3
V
Adjustment:
1. Adjust CAL:SETUP 22 for a DMM reading of 1.0606 mV
±
10 mV
Setup 39
500 mV Amplitude, Post Attenuators – Arbitrary
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range:
1
V
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination.
3. Configure the 3152B as follows:
Frequency: 1
kHz
Output:
On
Amplitude:
500
mV
Adjustment:
1.
Adjust CAL:SETUP 23 for a DMM reading of 176.7 mV
±
2 mV
Setup 40
100 mV Amplitude, Post Attenuators - Arbitrary
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
100
mV
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough