3152B User Manual
Publication No. 980935 Rev. A
Standard Waveforms
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
0.2 ms
Sampling Rate: 50 MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude:
1 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel 2.
4. Configure the function generator as follows:
Frequency 1
kHz
Run Mode:
Continuous
Waveform: Square
wave.
Amplitude: 2
V
5. Connect the function generator output connector to the 3152B
TRIG IN connector.
6. Configure model 3152B controls as follows:
Waveform:
Modulated
Modulation:
FM
Mod Run Mode: Triggered
Carrier Freq:
1 MHz
Mod Frequency: 10 kHz
Deviation: 500
kHz
Sync: On
Output: On
Test Procedure:
1. Verify triggered FM – standard waveforms operation on the
oscilloscope as follows:
Waveform:
Triggered sine waves
Mod Frequency: 10 kHz
Trigger Period: 1 ms
Max Deviation: 1.25 MHz
Min Deviation: 750 kHz
Test Results
Pass
Fail
FM Burst - Standard
Waveforms
Equipment: Oscilloscope, function generator
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
0.2 ms
Sampling Rate: 50 MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude:
1 V/div
2. Connect 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel
2.