3152B User Manual
Publication No. 980935 Rev. A
Figure 4-20, Counter/Timer Panel
Display
The Display Group has controls to select the display mode and to
select if the display shows measurement or gate time readings.
In normal mode, the counter is armed to receive signal at the trigger
input. When signal is sensed, the gate to the counter opens for
duration as programmed with the Gate Time parameter, processes
the result, displays the reading and continues with the same
process as long as the signal is available at the input.
In hold mode, the counter is armed to receive signal at the trigger
input. When signal is sensed, the gate to the counter opens for
duration as programmed with the Gate Time parameter processes
the result, displays and holds the reading until the next Reset/Arm
command.
To display and modify the gate time parameter, click on the Gate
Time LED and modify the gate time per your requirements. Gate
time rage is from 100
μ
s to 1 s. Normal counter/timer readings are
displayed when the Reading LED is selected.
Pulse Generator
The Pulse Generator panel (Figure 4-21) contains controls that
select the pulse function and adjusts the pulse parameters. The
pulses are generated digitally suing the arbitrary waveform
memory and digital computation and therefore, there are some
limitations to the minimum to maximum range that must be
observed. The pulse design limitations are given in Appendix A.
The various parameters that control the digital pulse generator
features are described below.