Publication No. 980935 Rev. A
3152B User Manual
EADS North America Test and Services
Programming Reference 5-63
<ampl>
10e-3 to
16
Numeric
5
Programs the amplitude of the standard sinc waveform
in units of Volts.
<offs>
-7.992 to
7.992
Numeric
0
Programs the offset of the standard sinc waveform in
units of Volts.
<cycles>
4 to 100
Integer
10
Programs the “0” crossing number of cycles of the
standard sinc waveform in units of percent.
Response
The 3152B returns the current frequency, amplitude, offset, and number of “0” crossing cycles settings as in
the following example: 1e6,5,0,10.
APPLy:GAUSsian <freq>,<ampl>,<offs>,<exp>(?)
HS Syntax: 0306D
<float>,<float,<float>,<integer>(?)
Description
This command changes the waveform function to standard gaussian and programs the frequency, amplitude,
offset, and exponent simultaneously. This command affects the output regardless of the current output
function. For example, if the 3152B is in FM mode, the 3152B will stop generating FM and will revert to the
standard Gaussian waveform with parametric values as specified by this command.
Parameters
Name
Range
Type
Default Description
<freq>
10e-3 to
100e6
Numeric
1e6
Programs the frequency of the standard gaussian
waveform in units of Hz.
<ampl>
10e-3 to
16
Numeric
5
Programs the amplitude of the standard gaussian
waveform in units of Volts.
<offs>
-7.992 to
7.992
Numeric
0
Programs the offset of the standard gaussian waveform
in units of Volts.
<exp>
4 to 100
Integer
10
Programs the exponent of the standard gaussian
waveform.
Response
The 3152B returns the current frequency, amplitude, offset and exponent settings as in the following example:
1e6,5,0,10.
APPLy:EXPonential <freq>,<ampl>,<offs>,<exp>(?)
HS Syntax: 0307D
<float>,<float,<float>,<integer>(?)
Description
This command changes the waveform function to standard exponential and programs the frequency,
amplitude, offset and exponent simultaneously. This command affects the output regardless of the current
output function. For example, if the 3152B is in FM mode, the 3152B will stop generating FM and will revert to
the standard exponential waveform with parametric values as specified by this command.
Parameters
Name
Range
Type
Default Description
<freq>
10e-3 to
100e6
Numeric
1e6
Programs the frequency of the standard exponential
waveform in units of Hz.
<ampl>
10e-3 to
Numeric
5
Programs the amplitude of the standard exponential
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com