3152B User Manual
Publication No. 980935 Rev. A
1. Configure the Oscilloscope as follows:
Termination: 50
Ω
, 20 dB feedthrough attenuator at the
oscilloscope input
Setup:
As required for the test
2. Connect the 3152B output to the oscilloscope input
3. Configure the 3152B as follows:
Frequency 1
MHz
Run Mode:
Triggered
Run Mode Src: As specified in Table 6-26
Waveform: Sine
wave
Amplitude:
2 V
Output: On
4. Configure the auxiliary 3152B as follows:
Frequency: 2
MHz
Waveform: Sine
wave
Run Mode:
Continuous
Trigger Output: As specified in Table 6-24
Output: On
Test Procedure
1. Set up the trigger output and trigger source as specified in
Table 6-24 and verify that the 3152B generates a single-
cycle, 2 MHz sine wave every 1
μ
s with every matched output
trigger and source settings
Table 6-24, Trigger Source Tests
Auxiliary 3152B
3152B Trigger
Oscilloscope
Trigger Output Setting
Source Setting
Reading
Pass
Fail
TTLT0 OFF TTLT1 ON
TTLT1
1
μ
s trig intervals
TTLT1 OFF TTLT2 ON
TTLT2
1
μ
s trig intervals
TTLT2 OFF TTLT3 ON
TTLT3
1
μ
s trig intervals
TTLT3 OFF TTLT4 ON
TTLT4
1
μ
s trig intervals
TTLT4 OFF TTLT5 ON
TTLT5
1
μ
s trig intervals
TTLT5 OFF TTLT6 ON
TTLT6
1
μ
s trig intervals
TTLT6 OFF TTLT7 ON
TTLT7
1
μ
s trig intervals
Sequence
operation
This tests the operation of the sequence generators. This also checks
the various sequence advance options.
Automatic Advance
Equipment: Counter
Preparation:
1. Configure the Counter as follows:
Function: TOTB
Measurement
2. Connect the counter channel B to the 3152B output
3. Configure the 3152B as follows:
6-16 Performance Checks
EADS North America Test and Services
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com