3152B User Manual
Publication No. 980935 Rev. A
Table 6-15, Sine Wave Spectral Purity, DDS Output Tests – at 5 V
3152B Freq
Reading
Spectrum Analyzer Settings
Settings Limits
Start
Stop
Reading
Pass
Fail
1 MHz
>45 dBc
100 kHz
10 MHz
5 MHz
>45 dBc
1 MHz
20 MHz
10 MHz
>35 dBc
1 MHz
100 MHz
50 MHz
>30 dBc
10 MHz
200 MHz
3. Change amplitude to 10 V and perform sine wave spectral purity,
DAC waveforms tests using Table 6-16.
Table 6-16, Sine Wave Spectral Purity, DDS Output Tests – at 10 V
3152B Freq
Reading
Spectrum Analyzer Settings
Settings Limits Start
Stop Reading
Pass
Fail
1 MHz
>35 dBc
100 kHz
10 MHz
5 MHz
>35 dBc
1 MHz
20 MHz
10 MHz
>30 dBc
1 MHz
100 MHz
50 MHz
>25 dBc
10 MHz
200 MHz
Sine Wave Flatness,
DAC Output
Equipment: Oscilloscope
Preparation:
1. Configure the Oscilloscope follows:
Termination: 50
Ω
, 20 dB feedthrough attenuator at the
oscilloscope input
Setup:
As required for the test
2. Connect 3152B output to the oscilloscope input
3. Configure the 3152B as follows:
Amplitude: 5
V
Output: On
Frequency:
Initially, 1 kHz then, as required by the test
Test Procedure
1. Adjust the vertical controls of the Oscilloscope to get 6 division
of display.
2. Perform Sine flatness, DAC waveforms tests using Table 6-17.
Table 6-17, Sine Wave Flatness, DAC Output Test – at 5 V
3152B Sine
Frequency
Error Limits
Oscilloscope Reading
Pass
Fail
1 kHz
6 Divisions
Reference
X
X
1 MHz
6 ±0.3 Divisions
10 MHz
6 ±0.3 Divisions
50 MHz
6 ±0.6 Divisions
100 MHz
6 ±0.6 Divisions
6-10 Performance Checks