Sampling Rate: 50 MS/s at least.
Trigger source: Channel 2, positive slope
Amplitude:
5 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel 2.
4. Configure the 3152B controls as follows:
Waveform:
Modulated
Modulation:
Amplitude Hopping
Hop Mode:
Fix
Dwell Time:
50
μ
s
Sync: On
Output: On
5. Using ArbConnection, open and prepare the Hop Table
composer and download the following table:
Amplitude
1
2
4
8
16
Test Procedure:
1. Verify 5 Hop operation on the oscilloscope as follows:
Waveform:
Amplitude steps, 50
μ
s fixed dwell time
Min Amp:
1 V
Max Amp:
16 V
Period: 250
μ
s
Test Results
Pass
Fail
Sweep
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
0.2 ms
Sampling Rate: 50 MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude:
1 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel 2
4. Configure model 3152B controls as follows:
Waveform:
Modulated
Modulation:
Sweep
Start Frequency: 1 MHz
Stop Frequency: 2 MHz
Sweep Time: 1 ms
Sweep Type: Linear
Sync: On
6-34 Performance Checks