3152B User Manual
Publication No. 980935 Rev. A
Preparation:
1. Configure the DMM as follows:
Function: ACV
Range:
1
V
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination.
3. Configure the 3152B as follows:
Frequency: 1
kHz
Output:
On
Amplitude:
500
mV
Offset:
250
mV
Adjustment:
1.
Adjust CAL:SETUP 19 for a DMM reading of 176.7 mV
±
2 mV
Setup 36
100 mV Amplitude, Amplifier Out - Arbitrary
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
100
mV
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination.
3. Configure the 3152B as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
100
mV
Offset:
250
mV
Adjustment:
1. Adjust CAL:SETUP 20 for a DMM reading of 35.35 mV
±
0.5 mV
Setup 37
50 mV Amplitude, Amplifier Out – Arbitrary
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
100
mV
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination.
3. Configure the 3152B as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
50
mV
Offset:
250
mV