Publication No. 980935 Rev. A
3152B User Manual
Adjustment
Procedures
Use the following procedures to calibrate the Model 3152B. The
following paragraphs show how to set up the instrument for calibration
and what the acceptable calibration limits are.
Calibration is done with the covers closed and the 3152B installed in a
VXI chassis. Register-based or message-based interfaces can be
used and the procedure is the same for both options. The calibration
procedure requires that the ArbConnection utility be installed and
interfaced to the instrument.
Calibration is performed from the Calibration Panel in ArbConnection.
To invoke this panel, one requires a password that is available to
service centers only. Contact EADS North America Test and Services
Customer Support to obtain your calibration password. Use the
following procedure to calibrate the generator:
1. Invoke
ArbConnection.
2. Click on the Calibration tab on the Panels bar.
3. When the Enter Password dialog box appears (Figure 7-1), type
your User Name and Password, and then click OK.
The Calibration Panel shown in Figure 7-2 will appear.
Figure 7-1, Enter Password Dialog Box
EADS North America Test and Services
Adjustments and Firmware Update 7-3