Installation
ZW User's Manual
2
Install
a
tion an
d
C
onn
ecti
ons
37
Installing on bottom
The following describes how to attach the Sensor Controller on its bottom.
Connecting Calibration ROM
To connect the calibration ROM to the Sensor Controller, follow the steps below:
• Before connecting or disconnecting the calibration ROM, make sure that the Sensor Controller's power supply is
turned OFF. The Sensor Controller may break down if the calibration ROM is connected while the power is ON.
• Use with the Calibration ROM always connected. If the Calibration ROM is not connected, an error is displayed.
• Only a calibration ROM and Sensor Head with a same serial number are compatible. When connecting a calibration
ROM with a Sensor Head, make sure that they have a same serial number, or measurement cannot be performed
correctly.
1
Drill four installation holes on the base.
For the location installation holes, see the external
dimensions.
10-1 Specifications and External Dimensions
2
Tighten four M4 screws to fix the Sensor Controller
on the base.
Tightening torque: 1.2 N • m
Sensor Controller thread depth: 6 mm
1
Connect the calibration ROM to the ROM connector
on the Sensor Controller.
2
Fix the calibration ROM by tightening the supplied
M2 screw.
Tightening torque: 0.15 N • m max.
Important
M4 scre
w
× 4
M2 screw
Calibration ROM
ROM connector
Important
Summary of Contents for ZW-C1*AT
Page 17: ...ZW User s Manual 1 15 ...
Page 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Page 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Page 120: ...Setting the System 118 ZW User s Manual MEMO ...
Page 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Page 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Page 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Page 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Page 449: ......