Setting the Output Conditions
106
ZW User's Manual
Setting the Offset
Set this item to set the reference value by a zero reset to a value other than zero.
Explorer pane
: [Bank] | [(Bank Data Name)] (double click)
Edit pane
: [Task Settings] icon (
)
Task Settings window : [Zero reset]
Item
Setting item
Setting value
Description
Zero reset
Offset
-999.999999 to 999.999999 [mm]
(default value: 0)
Set the offset value.
1
Set [Status] to ON.
Offset can be set only when [Status] is set to ON.
2
Enter an offset value at [Zero reset] - [Offset].
The offset can also be set by the operating keys on the Sensor Controller.
Important
Note