
Setting the Output Conditions
ZW User's Manual
99
4
Settings for
Func
tion
Explorer pane
: [Bank] | [(Bank Data Name)] (double click)
Edit pane
: [Task Settings] icon (
)
Task Settings window : [Hold]
Hold
Hold mode
Auto Peak to
Peak
Holds the difference between the maximum and minimum values of the
measurement result. The output changes every time the maximum or the
minimum value is updated.
Average
Holds the "average of the measured value" of the sampling period. Output
changes after sampling, and this state is held until the next sampling ends.
Sampling
Holds the measured value the moment that control enters the sampling
period. Output changes at the start of sampling, and this state is held until the
next sampling is started.
Through
(default value)
The hold function is not used. The measured value is always output.
1
Select the hold mode from [Hold mode].
Setting items for required parameters are
displayed depending on the selected hold
mode.
The hold mode can also be set by the operating keys on the Sensor Controller.
Item
Setting item
Setting value
Description
0
0
Current
measured value
Output
(Max.value - Min.value)
Current
measured value
Output
(Average)
Sampling period
Current
measured value
Output
Sampling period
Note
Summary of Contents for ZW-C1*AT
Page 17: ...ZW User s Manual 1 15 ...
Page 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Page 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Page 120: ...Setting the System 118 ZW User s Manual MEMO ...
Page 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Page 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Page 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Page 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Page 449: ......