Connecting Parallel I/O
346
ZW User's Manual
Setting for Internal Logging
Setting LOGGING save count and LOGGING save intervals
Settings for Internal Logging p.151
As an example, here is an explanation of the procedure for setting save intervals to "1" and save count to "100".
Operating procedure
Steps
Key operation
Display
Description
1
Press and hold the
key for two seconds to enter the FUN mode.
2
Press the
/
keys to select "I/O" and press the
key.
3
Press the
/
keys to select "PAR.LOG" and press the
key.
4
Select the save intervals.
Press the
/
keys to select "INTRVL" and press the
key.
5
Set the save intervals.
The current setting value is displayed on the sub-display.
Press
key to enter the editing mode, and the sub-display blinks.
To edit numerical values, use the
/
/
/
keys.
In this example, input "1", then press the
key.
6
Press the
key to return to the previous menu.
7
Select the save count.
Press the
/
keys to select "COUNT" and press the
key.
8
Set the save count.
The current setting value is displayed on the sub-display.
Press
key to enter the editing mode, and the sub-display blinks.
To edit numerical values, use the
/
/
/
keys.
In this example, input "100", then press the
key.
Summary of Contents for ZW-C1*AT
Page 17: ...ZW User s Manual 1 15 ...
Page 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Page 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Page 120: ...Setting the System 118 ZW User s Manual MEMO ...
Page 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Page 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Page 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Page 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Page 449: ......