No-protocol Connection
230
ZW User's Manual
Setting Communications Specifications (RS-232C Communications)
Setting RS-232C communications on the sensor body
Set RS-232C communications on the displacement sensor body.
Explorer pane
: [Device Group] | [(Sensor Name)] | [System] |
[System Data] (double-click)
Edit pane
: [RS-232C Communications Settings] icon (
)
Item
Description
Range
Baud rate
Sets the data transfer speed.
9600bps, 19200bps, 38400bps (default value),
57600bps, 115200bps
Data length
Sets the data length.
8 bits (default value), 7 bits
Parity
Sets the parity bit (error detection sign).
None (default value), odd, even
Stop bit
Sets the stop bit.
1bit (default), 2bit
Delimiter
Sets the delimiter (data delimiter).
CR (default), LF, CR+LF
CS/RS
Sets the flow control.
OFF(default value)/ON
• With the ZW series, communication cannot be established under the following condition.
Data length: 7-bit and Parity: None
• The RS-232C communication specifications can also be set with key operations on the Sensor Controller.
9-15 Connecting by No-protocol Communications p.352
Note
Summary of Contents for ZW-C1*AT
Page 17: ...ZW User s Manual 1 15 ...
Page 54: ...Installing the Sysmac Studio 52 ZW User s Manual MEMO ...
Page 74: ...Operating with Sensor Controller 72 ZW User s Manual MEMO ...
Page 120: ...Setting the System 118 ZW User s Manual MEMO ...
Page 138: ...Controll input signal with PC tool 136 ZW User s Manual MEMO ...
Page 262: ...No protocol Connection 260 ZW User s Manual MEMO ...
Page 271: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 270 8 2 Troubleshooting 284 ...
Page 358: ...Connecting by No protocol Communications 356 ZW User s Manual MEMO ...
Page 449: ......