Figure 62. RV Capture Threshold trend detail
6.14 Automatic device status monitoring
The device automatically and continuously monitors for adequate charge time performance,
electrical reset, and disabled therapy conditions. During each interrogation, the device
reports detected conditions that require attention as device status indicator warnings and
then displays these warnings on the programmer screen. A device status indicator warning
is displayed as a pop-up window on the programmer screen and is displayed also in the
Observations box on the Quick Look II screen. A specific procedure about how to respond
to the device status indicator warning for electrical reset is provided in Section 6.14.2, “How
to respond to the device status indicator warning for electrical reset”, page 196.
Caution: The device status indicators are important. Please inform your Medtronic
representative if any of the indicators are displayed on the programmer screen after
interrogating a device.
To clear the displayed status indicator, select [Clear] in the pop-up window that displays the
device status indicator warning.
6.14.1 Definitions of device status indicator warnings
Warning - Charge Circuit Timeout – Indicates that the charging period has exceeded
30 s. The charge circuit is still active. Inform a Medtronic representative if this device status
Medtronic
PROTECTA™ XT DR D314DRG
194
Clinician Manual