Check the Arrhythmia Episodes window for High Rate-NS episodes. A High Rate-NS
episode has an average ventricular rate greater than 273 bpm (less than 220 ms). If there
are 2 or more High Rate-NS episodes, lead noise oversensing may have occurred. Review
the stored EGM to determine the cause of the oversensing.
Figure 32. High Rate-NS episode record
1 Average rate information
6.4.5.6 Lead Trends
Select Data icon
⇒ Device/Lead Diagnostics
⇒ Lead Impedance Trends
Check the Lead Impedance Trends for a sudden change in the RV Pacing Impedance
measurement. If at least 1 measurement is greater than 175% of the baseline value or is
less than 50% of the baseline value, then the lead impedance should be considered
abnormal.