Note: Conducted AF Response operation is suspended during automatic tachyarrhythmia
therapies, system tests, EP study inductions, manual therapies, and emergency fixed burst,
cardioversion, and defibrillation. Conducted AF Response operation is not suspended
during an impedance test or a Charge/Dump Test.
7.18.3 Programming considerations for Conducted AF Response
Maximum Rate – Increases to the pacing rate caused by Conducted AF Response are
limited by the programmed Maximum Rate.
Response Level value – A higher Response Level value results in a higher percentage of
ventricular pacing and faster alignment with the patient’s own ventricular response rate.
DDD or DDDR mode – Conducted AF Response operates only in nontracking modes.
Therefore, when the device is programmed to DDD or DDDR mode, Conducted AF
Response operates only during a mode switch to DDIR mode. Mode Switch must be
programmed to On to program Conducted AF Response to On.
Conducted AF Response and VRS – In DDIR and VVIR modes, Conducted AF Response
and VRS cannot be programmed to On at the same time.
7.18.4 Programming Conducted AF Response
Select Params icon
⇒ Pacing…
⇒ Arrhythmia/Post Shock…
▷ Conducted AF Response <On>
⇒ Additional V Settings…
▷ Response Level
▷ Maximum Rate
7.18.5 Evaluation of Conducted AF Response
The device reports ventricular rates during AT/AF episodes in the Rate Histogram and
Cardiac Compass Trends reports. This information may help you decide which Response
Level and Maximum Rate values to select for Conducted AF Response.
Select Reports icon
⇒ Available Reports…
⇒ Rate Histograms