12. Observe the Live Rhythm Monitor for proper detection, therapy, and post-shock
sensing.
13. To review the stored data for the induced episode, select [Retrieve Data…]. To view
more details, print a Last VT/VF with EGM report, or select Data > Clinical
Diagnostics > Arrhythmia Episodes to view the data on the programmer.
14. Select [Adjust Permanent…] to program a new VF Therapy Rx1 energy level or to
change the Pathway, if desired.
15. Wait until the on-screen timer reaches 5 min, then repeat Step 11 through Step 15 as
needed.
16. Before closing the pocket, select the Params icon and program VF Detection, FVT
Detection, and VT Detection to Off.
5.6 Positioning and securing the device
Caution: If an SVC electrode is not implanted, make sure the pin plug provided with the
device is inserted into the SVC port to prevent electrical leakage.
Caution: Program tachyarrhythmia detection to Off or Monitor to avoid inappropriate
detection or therapy delivery while closing the surgical pocket.
Note: Implant the device within 5 cm (2 in) of the surface of the skin to optimize post-implant
ambulatory monitoring. The side of the device engraved with the Medtronic logo should face
toward the skin so it is easier for the patient to hear the alert tones.
5.6.1 How to position and secure the device
1. Verify that each lead connector pin or pin plug is fully inserted into the connector port
and that all setscrews are tight.
2. To prevent twisting of the lead body, rotate the device to loosely wrap the excess lead
length (see Figure 25). Do not kink the lead body.
Medtronic
PROTECTA™ XT DR D314DRG
120
Clinician Manual