Figure 33. RV Lead Impedance Trend data
1 Baseline value
2 Abnormal impedance measurements
6.5 Viewing long-term clinical trends with the Cardiac
Compass Report
An analysis of clinical information collected over a long term can help you to follow changes
in a patient’s condition and correlate these changes with variations in device programming,
medication, patient activity, or symptoms.
The Cardiac Compass Report provides a picture of the patient’s condition over the last 14
months. Graphs show trends in the frequency of arrhythmias, the amount of physical activity,
heart rates, and device therapies. Dates and event annotations allow you to correlate trends
from different graphs. The report can also help you to assess whether device therapies or
antiarrhythmic drugs are effective.
Cardiac Compass trend data is available only as a printed report.
The Cardiac Compass Report is based on data and measurements collected daily. Data
storage for the Cardiac Compass Report is automatic. No setup is required. The device
begins storing data after the device is implanted. Each day thereafter, the device stores a
set of Cardiac Compass trend data. Storage continues until the 14-month storage capacity
is filled. At that point, the oldest stored data is overwritten with new data.
Medtronic
PROTECTA™ XT DR D314DRG
148
Clinician Manual