![Medtronic PROTECTA XT DR D314DRG Clinician Manual Download Page 229](http://html.mh-extra.com/html/medtronic/protecta-xt-dr-d314drg/protecta-xt-dr-d314drg_clinician-manual_1763148229.webp)
Figure 81. Pacing mode on the status bar
7.3.5.2 Quick Look II screen
Select Data icon
⇒ Quick Look II
The Quick Look II screen shows the percentages of atrial and ventricular pacing since the
last follow-up appointment. The Quick Look II screen also reports if the device is
programmed to an MVP mode. If the present programmed pacing mode is AAIR<=>DDDR
or AAI<=>DDD, the message “MVP On” appears on the Quick Look II screen. Otherwise,
the screen displays “MVP Off.”
For detailed information about viewing and interpreting all of the information available from
the Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,
7.3.5.3 Cardiac Compass Report
Select Data icon
⇒ Clinical Diagnostics
⇒ Cardiac Compass Trends (Report Only)
The % Pacing/day trend on the Cardiac Compass Report shows changes in the frequency
of atrial and ventricular pacing over time.
Figure 82. Cardiac Compass Report
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
229