established and the programming head is positioned over the device. The device stops
these transmissions when you lift the programming head, unless the Holter Telemetry
feature is programmed to On. If Holter Telemetry is programmed to On, the device transmits
Marker Channel and supplementary marker data regardless of the position of the
programming head.
Device longevity and wireless telemetry – In typical patient session and device operation
scenarios, wireless telemetry has no significant effect on device longevity.
4.2.3 How to interrogate the device during the session
At the start of the patient session the programmer interrogates the device. You can manually
interrogate the device at any time during the patient session by performing the following
steps:
1. Select [Interrogate…] from the Command bar. In a nonwireless session, you may also
interrogate the device by pressing the “I” button on the programming head.
2. To gather information collected since the last patient session, select the Since last
session option from the interrogation window. If you want to gather all of the information
from the device, select the All option.
3. Select [Start].
Note: You cannot manually interrogate the device during an Emergency. You must exit an
Emergency before you can manually interrogate the device.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
55