11 The [Print…] button prints the on-screen frozen strip.
12 The [Close] button closes the frozen strip viewing window.
4.11.4 Recalling waveform strips
Before ending the patient session, you can recall any waveform strip collected and saved
during the session in order to view, adjust, and print the waveform strip.
4.11.4.1 How to recall a waveform strip
1. Select [Strips…] in the tool palette or in the strip viewer.
2. Select a strip to view.
3. Select [Open]. The strip viewer displays the selected strip.
4.12 Expediting follow-up sessions with Leadless ECG
An analysis of a patient’s real-time ECG signal is an important part of most follow-up
assessments. Connecting surface leads to the patient and acquiring an acceptable ECG
signal can be a time consuming part of a follow-up session. Additionally, connecting surface
leads to a patient requires that the patient be present in the clinic.
Medtronic
PROTECTA™ XT DR D314DRG
92
Clinician Manual