Table 53. Lead/Device Integrity Alerts (continued)
Parameter
Programmable values
Shipped Reset
RV Pacing Impedance Ena-
Off; On
On
Off
RV Defibrillation Impedance
Enable
Off; On
On
Off
SVC Defibrillation Impe-
dance Enable
Off; On
On
Off
Shared (Device Tone and Patient Home Monitor)
A. Pacing Impedance Less
than
200 ; 300; 400; 500 Ω
200 Ω
200 Ω
A. Pacing Impedance
Greater than
1000; 1500; 2000; 3000 Ω
3000 Ω
3000 Ω
RV Pacing Impedance Less
than
200 ; 300; 400; 500 Ω
200 Ω
200 Ω
RV Pacing Impedance
Greater than
1000; 1500; 2000; 3000 Ω
3000 Ω
3000 Ω
RV Defibrillation Impedance
Less than
20 ; 30; 40; 50 Ω
20 Ω
20 Ω
RV Defibrillation Impedance
Greater than
100; 130; 160; 200 Ω
200 Ω
200 Ω
SVC Defibrillation Impe-
dance Less than
20 ; 30; 40; 50 Ω
20 Ω
20 Ω
SVC Defibrillation Impe-
dance Greater than
100; 130; 160; 200 Ω
200 Ω
200 Ω
Low Battery Voltage RRT…
Device Tone
Alert Enable - Urgency
Off; On-Low; On-High
On-High Off
Patient Home Monitor
Alert Enable
Off; On
On
Off
Excessive Charge Time EOS…
Device Tone
Alert Enable - Urgency
Off; On-Low; On-High
On-High Off
Patient Home Monitor
Alert Enable
Off; On
On
Off
VF Detection Off, 3+ VF or 3+ FVT Rx Off.
Device Tone
Alert Enable
Off; On-High
On-High On-High
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
473