If you save data to a diskette that is corrupt or is not IBM-formatted, the programmer may
become unresponsive. If this occurs, remove the diskette, and turn the programmer off and
then on again. Normal operation should resume. Please inform your Medtronic
representative of this occurrence.
4.13.1.2 Considerations for saving device data to a diskette
Emergency functions while saving – During the save operation, the [Emergency] button
remains displayed, and all Emergency functions are available. If a disk error occurs during
a save, there may be a delay in initiating the Emergency screens. Therefore, it is suggested
that you not save to disk during EP studies or when it is possible that Emergency functions
will be needed immediately. If an Emergency function is used during a save operation, the
device aborts the save operation.
Interrogate first – Interrogate the device before saving data to a diskette because the
programmer saves only the data it has interrogated. If you want to save a record of all the
information from the device, select the All option from the interrogation window. Selecting
the All option provides more data for analysis if an issue needs to be investigated.
4.13.1.3 How to save device data to a diskette
1. Select [Interrogate…] to interrogate the device.
2. Select Session > Save to Disk….
3. Insert a diskette into the programmer diskette drive.
4. Select [Save].
You also have the option to Save to Disk when you select [End Session…].
4.13.2 Retrieving device data from a diskette
When the programmer has read the data that was saved during a patient session, it presents
the information in a read-only view. In the read-only view, the data is presented in a slightly
different way than what is seen in a live session. No Live Rhythm Monitor is displayed
because this is not a live session. Instead, the Live Rhythm Monitor is replaced with the
device model and the words Read From Disk. While in the Read From Disk application, the
programmer allows you to view the saved data, print reports, and display all programmed
parameter values.
Medtronic
PROTECTA™ XT DR D314DRG
94
Clinician Manual