9.2.2.4 Ramp+ pacing therapy
The programmable parameter Initial # Pulses sets the number of pulses in the first Ramp+
sequence. Ramp+ pacing intervals are controlled by the programmable parameters R-S1
Interval=(%RR), S1S2(Ramp+)=(%RR), and S2SN(Ramp+)=(%RR).
The pulses in the Ramp+ sequence are delivered at progressively shorter pacing intervals,
each determined as a specific percentage of the ventricular tachycardia cycle length. The
first pulse of each Ramp+ sequence is delivered at a pacing interval using the R-S1
Interval=(%RR) percentage. The pacing interval for the second pulse in the sequence is
determined by the S1S2(Ramp+)=(%RR) percentage. Any remaining pulses in the
sequence are delivered at a pacing interval using the S2SN(Ramp+)=(%RR) percentage.
If the ventricular tachycardia is redetected, the device applies the programmed percentages
to the new ventricular tachycardia cycle length to determine the pacing intervals for the next
Ramp+ sequence. Each sequence adds one pacing pulse.
Note: The Ramp+ pacing therapy is delivered in the VOO pacing mode.
In the example of Ramp+ pacing operation in Figure 192, two Ramp+ therapy sequences
are delivered. The second therapy sequence terminates the VT episode.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
373