Note: Similar programming steps apply for Ramp therapies for Fast AT/AF episodes.
9.5.4.3 Programming 50 Hz Burst pacing therapy
Select Params icon
⇒ AT/AF | Therapies…
⇒ AT/AF Rx | Anti-Tachy Pacing (ATP)…
▷ AT/AF Rx Status <On>
▷ Therapy Type <50 Hz>
▷ 50 Hz Burst Duration
▷ # Sequences
Note: Similar programming steps apply for 50 Hz Burst therapies for Fast AT/AF episodes.
9.5.4.4 Programming Shared A. ATP therapies
Select Params icon
⇒ AT/AF | Therapies…
⇒ AT/AF Rx | Anti-Tachy Pacing (ATP)…
▷ A-A Minimum ATP Interval
▷ A. Pacing Amplitude
▷ A. Pacing Pulse Width
▷ VVI/VOO Backup Pacing
▷ VVI/VOO Backup Pacing Rate
9.5.5 Evaluation of atrial ATP therapies
9.5.5.1 The Quick Look II screen
Select Data icon
⇒ Quick Look II
Treated AT/AF episodes – This section includes a count of treated AT/AF episodes. You
can select the Treated [>>] button to view the data for treated episodes.
Quick Look II observations – The Quick Look II observations are based on an analysis
of interrogated data since the last session and programmed parameters. If related
information about an observation is available, you can select the observation and then select
the Observations [>>] button to view related information.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
405