Figure 179. Episode EGM of a T-wave oversensing episode
3
1 Marker Channel shows each ventricular event is counted twice.
2 TW annotated upon initial detection of T-wave oversensing.
3 TW annotated to indicate T-wave oversensing on a per-event basis.
8.9.5.3 Episode Text
Select Data icon
⇒ Clinical Diagnostics
⇒ Arrhythmia Episodes
⇒ Text
For T-wave oversensing episodes, the Episode Text screen lists the following types of
information:
●
an episode summary for V.Oversensing-TWave including:
– Duration
– A/V Max Rate
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
353