4.1.1.5 How to maintain reliable telemetry
You can expect reliable wireless telemetry between the implanted device and the
programmer in a typical examination room or operating room. If the device and programmer
are in the presence of other electrical equipment, the system is designed to maintain
effective communication at distances up to 2 m (6.5 feet) between the device and the
programmer. The system should not interfere with other electronic equipment in the area.
If you are having trouble maintaining consistent, reliable telemetry between a patient’s
implanted device and the programmer, take one or more of the following actions to increase
the number of illuminated lights on the telemetry status indicator:
●
Adjust the angle of the programmer screen. The telemetry antenna is part of the
programmer display screen structure, and slight movements of the screen may improve
the telemetry link.
●
Change the position of the programmer so that the space between the programmer
screen and the patient is relatively free of obstruction. The optimal position for the
programmer is between the patient and you, so that you are facing the screen and the
patient is beyond the screen facing both you and the programmer.
●
Shorten the distance between the programmer and the patient.
●
Signal strength may be stronger with the device placed in the patient than if the device
is in the packaging.
●
Remove any sources of electromagnetic interference (EMI) that may be affecting the
telemetry signal.
If a session is accidently disrupted by electrical interference, the programmer attempts to
reestablish communication with the device for the next 5 min. If communication between
the device and programmer is not reestablished during that period, you must use the
Conexus Activator or programming head to reactivate wireless telemetry in the device to
resume or to start a new session.
Note: If you want to use Holter Telemetry to transmit EGM and Marker Channel data during
a Conexus telemetry session, you must first activate Standby mode.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
49