![Medtronic PROTECTA XT DR D314DRG Clinician Manual Download Page 142](http://html.mh-extra.com/html/medtronic/protecta-xt-dr-d314drg/protecta-xt-dr-d314drg_clinician-manual_1763148142.webp)
If the possibility of an impending RV lead fracture is indicated by the data, the RV Lead
Integrity Alert feature can trigger a Medtronic CareAlert Notification and an alert tone to warn
the patient. In addition, the RV Lead Integrity Alert feature automatically adjusts
tachyarrhythmia detection settings and diagnostic settings to avoid the delivery of an
inappropriate shock.
6.4.2 Operation of RV Lead Integrity Alert
Figure 30. Operation of RV Lead Integrity Alert
RV pacing lead
impedance
Response
Sensing Integrity Counter
High Rate-NS episodes
60 days
2 of 3 criteria
are met
CareAlert
notification
Changes to EGM
storage
Adjustments to
tachyarrhythmia
detection
The device continually monitors for a potential RV lead fracture using lead impedance
measurements, the Sensing Integrity Counter, and High Rate-NS episode data. It identifies
a potential lead fracture if at least 2 of the following criteria have been met within the past
60 days:
●
An RV pacing lead impedance measurement is less than 50% or greater than 175% of
the baseline impedance. The baseline measurement is the median of the previous 13
daily measurements.
●
The ventricular Sensing Integrity Counter is incremented by at least 30 within a period
of 3 consecutive days or less.
Note: The Sensing Integrity Counter total displayed on the Battery and Lead
Measurements screen is the number of short ventricular intervals that occurred since
the last patient session. Therefore, the total could exceed 30 without satisfying the alert
criteria if the total was reached during a period of more than 3 consecutive days.
●
The device senses 2 High Rate-NS episodes with a 4-beat average R-R interval of less
than 220 ms.